Magnetoacoustic tomographic imaging of electrical impedance with magnetic induction.

Mark McIntosh,'s picture
PubMed URL: 
http://www.ncbi.nlm.nih.gov/pubmed/19169372
Author: 
He B
Author List: 
Xia R
Li X
He B
Journal: 
Appl Phys Lett
PubMed ID: 
19169372
Pagination: 
83903
Volume: 
91
Issue: 
8
Abstract: 
Magnetoacoustic tomography with magnetic induction (MAT-MI) is a recently introduced method for imaging tissue electrical impedance properties by integrating magnetic induction and ultrasound measurements. In the present study, we have developed a focused cylindrical scanning mode MAT-MI system and the corresponding reconstruction algorithms. Using this system, we demonstrated 3-dimensional MAT-MI imaging in a physical phantom, with cylindrical scanning combined with ultrasound focusing, and the ability of MAT-MI in imaging electrical conductivity properties of biological tissue.
Published Date: 
August, 2007